LIVANOVA USA AORTIC ROOT CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number AR-11112
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Livanova usa received a report that surgeon was using the ar-11112 and was got stubbed using the steel introducer needle.There is no report of any patient injury.
 
Additional Manufacturer Narrative
A.1.-a.5.Patient information was not provided.H10: livanova usa manufactures the aortic root cannula.The incident occurred in usa.Complained cannula has been requested for investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event or Problem Description
See initial report.
 
Additional Manufacturer Narrative
H10: the complained unit was returned to livanova for investigation, along with 9 other unused units from the same lot that were refused by customer.Visual inspection concluded that the device was manufactured properly per manufacturing procedures.When the steel introducer needle was being pulled out of the cannula, the luer was slightly more difficult to twist, however that may have been due to the fluid that dried on the device causing it to stick.Once the luer was untwisted, there were no issues removing the needle and it functioned properly.Dimensional analysis was performed on critical dimensions that could contribute to the needle being more difficult than usual to remove, on the tubing and on the needle.All dimensions were found to be within specifications.Visual inspection was performed also on the 9 unused units and these were manufactured properly per manufacturing procedures.All luers were untwisted with minimal effort and there were no issues removing the needle.The device history record (dhr) review highlighted that the lot whose claimed product belongs to was released as conform according to specifications.No similar complaints have been recorded on this product lot out of 300 manufactured units nor part number in the last 5 years.Therefore, the event is isolated.Upon reviewing product change history, no changes that would contribute to the reported complaint were identified.Based on all the above, it cannot be ruled out that the reported event could have been due to the needle luer that may have been assembled too tightly to the female connector of cannula tubing during manufacturing, making needle withdrawal harder for the surgeon.To prevent re-occurrence, the manufacturing personnel has been involved in a dedicated training meeting to discuss the specific event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC ROOT CANNULA
Common Device Name
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA USA
arvada CO 80004
Manufacturer (Section G)
LIVANOVA USA INC.
14401 w 65th way
arvada CO 80004
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18433430
Report Number1718850-2024-00021
Device Sequence Number15212138
Product Code DWF
UDI-Device Identifier00803622103049
UDI-Public(01)00803622103049(240)AR-11112(17)260803(10)2321600074
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K972503
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/03/2026
Device Catalogue NumberAR-11112
Device Lot Number2321600074
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2023
Supplement Date Received by Manufacturer03/30/2024
Initial Report FDA Received Date01/03/2024
Supplement Report FDA Received Date04/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-