| Model Number |
60-01-04 |
| Medical Device Problem Code |
Pumping Stopped (1503)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the centrifugal pump system with tubing clamp.The incident occurred in coventry, united kingdom.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova deutschland received a report that a centrifugal pump system with tubing clamp gave motor control failure error message.There was no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported error code can be due to a disconnection between the motor control board (hms) and the motor.A livanova field service representative was dispatched to the facility and confirmed the reported issue.The unit was shipped to livanova deutschland facility for repair activity that has not been performed yet.The specific root cause cannot be determined.However, considering investigation's results of similar events, the most likely root cause of the reported issue could be assigned to a faulty connection of the motor on the motor control board (hms) board.Failure of electronic boards can be related to multiple and not deterministic factors such as exposure to heat, dust and moisture, accidental impacts (drops and falls), and power overloads/surges but also to variability of micro sub-components.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|