| Catalog Number |
AA6106 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
Urinary Retention (2119)
|
| Date of Event |
11/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the available information bladder globes on repeating catheters.
|
| |
|
Event or Problem Description
|
|
According to the available information bladder globes on repeating catheters.
|
| |
|
Additional Manufacturer Narrative
|
|
After receiving this complaint, we searched for other complaints and we found none regarding the lot number 9261039 with the same defect.This product was made by our subcontractor which was informed about this issue.In february we received subcontractor's investigation which revealed no anomaly recorded during production.The potential root cause could be a raw material's balloon.A similar case study was performed based on same item number aa6106, same defect: drainage issue over last four years: one similar case was found.Without the sample we cannot do more than the documentary investigation which revealed no anomaly in relation to the describe defect on this complaint.
|
| |
|