• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP Back to Search Results
Catalog Number 1801-58
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported when the catheter was prepared and the air inside the balloon was removed with normal saline, the balloon easily ruptured.The issue occurred with 2 catheters.Another catheter of the same size was prepared and the procedure was successfully completed using the replacement.No injury was reported to the patient.Note: this is report 1 of 2 related to the first catheter used in the event.Report 1220948-2023-00215 was submitted for the second device involved in this event.
 
Additional Manufacturer Narrative
The device was received for investigation.The reported problem was confirmed.The balloon was observed to be ruptured.While the reported event was confirmed, the exact root cause of the balloon rupture could not be determined.The lot history record for the device was reviewed, no issues were found during manufacturing or packaging that would cause or contribute to the reported event.The ifu states: complications associated with the use of embolectomy catheters include, but are not limited to: systemic infection, local hematomas, intimal disruptions, arterial dissection, perforation and rupture, hemorrhage, arterial thrombosis, distal emboli of blood clots or arteriosclerotic plaque, air embolus, aneurysms, arterial spasms, arteriovenous fistula formations, and balloon rupture or tip separation with fragmentation and distal embolization.Note: this is report 1 of 2 related to the first catheter used in the event.Report 1220948-2023-00215 was submitted for the second device involved in this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP
Common Device Name
EMBOLECTOMY CATHETER - REGULAR TIP
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC
63 second ave
burlington MA 01803
Manufacturer Contact
peter song
63 second ave
burlington, MA 01803
7814251683
MDR Report Key18436565
Report Number1220948-2023-00214
Device Sequence Number7781559
Product Code DXE
UDI-Device Identifier00840663100941
UDI-Public(01)00840663100941
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K992368
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1801-58
Device Lot NumberNSE2455
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2023
Initial Report FDA Received Date01/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-