| Catalog Number |
INSC-7-230-S |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The system preparation section of the instructions for use includes the following: "uncoil device.Verify smooth handle operation and clip action.Open clip by gently moving handle spool distally (away from handle thumb ring).Once clip is fully open, do not continue advancing handle spool as clip may prematurely detach from catheter.Close clip by moving handle spool proximally until clip is fully closed.Caution: do not continue to pull handle spool beyond tactile resistance as this may prematurely deploy clip." the instructions for use also states: "with clip closed and without holding handle spool, advance device in small increments into accessory channel of gastroscope or colonoscope.Caution: holding handle spool during clip advancement may prematurely deploy clip." failure to follow the instructions above can result in damage to the device which may lead to premature clip deployment.Prior to distribution, all instinct endoscopic hemoclips are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During the upper digestive endoscopy procedure with mucosectomy, the physician used a cook instinct endoscopic hemoclip.It was reported that when positioning the clip, it fired deceptively [prematurely deployed in the unknown position].The clip was removed with biopsy forceps.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction follow up report is being submitted to cancel the initial report submitted relating to this event.The initial report was that during the procedure, when positioning the clip, it fired deceptively.Cook interpreted that information as the clip prematurely deploying in an unknown position which is a reportable event.The facility responded to cook's request for additional information stating that the clip had deployed in the closed position.This changes the event to premature deployment in the closed position which has not historically caused or contributed to a serious injury or death.Based on further quality and medical evaluation, this incident no longer meets the reporting criteria of an fda mdr report.
|
| |
|
Event or Problem Description
|
|
This follow up emdr is being sent to cancel the initial emdr sent on this event.
|
| |
|