• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 90 DEG CURVE STRAIGHT, QUICKPASS LASSO; ORTHOPEDIC MANUAL SURG INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 90 DEG CURVE STRAIGHT, QUICKPASS LASSO; ORTHOPEDIC MANUAL SURG INSTR Back to Search Results
Model Number 90 DEG CURVE STRAIGHT, QUICKPASS LASSO
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 12/06/2023, it was reported by an arthrex subsidiary employee that an ar-6068-90 quick pass lasso tip of the shaft broke during use.The tip was removed and the case was completed using a new product.This occurred during use in a case with no patient effect.
 
Additional Manufacturer Narrative
Complaint allegation is confirmed.Upon visual inspection, it was noted that the tip was broken off and the end of the shaft was cut off.Functional testing was not performed due to the damage to the device.The most likely cause for the reported failure can be attributed misaligned insertion; prying/leveraging the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
90 DEG CURVE STRAIGHT, QUICKPASS LASSO
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18437688
Report Number1220246-2024-00058
Device Sequence Number7155411
Product Code LXH
UDI-Device Identifier00888867038943
UDI-Public00888867038943
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number90 DEG CURVE STRAIGHT, QUICKPASS LASSO
Device Catalogue NumberAR-6068-90
Device Lot Number3159132362
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2023
Supplement Date Received by Manufacturer07/16/2024
Initial Report FDA Received Date01/03/2024
Supplement Report FDA Received Date08/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-