| Model Number |
90 DEG CURVE STRAIGHT, QUICKPASS LASSO |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 12/06/2023, it was reported by an arthrex subsidiary employee that an ar-6068-90 quick pass lasso tip of the shaft broke during use.The tip was removed and the case was completed using a new product.This occurred during use in a case with no patient effect.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.Upon visual inspection, it was noted that the tip was broken off and the end of the shaft was cut off.Functional testing was not performed due to the damage to the device.The most likely cause for the reported failure can be attributed misaligned insertion; prying/leveraging the device during insertion.
|
| |
|