• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACCUSTICK II; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACCUSTICK II; INTRODUCER, CATHETER Back to Search Results
Model Number 35615
Medical Device Problem Codes Defective Device (2588); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the markerband was dislodged.The target lesion was located in the kidney.An accustick ii w/o guidewire was selected for use.After withdrawing the device from the patient, it was noted that the silver markerband was found in the upper part of the dilator and not at the end of the introducer.The procedure was completed with this device.No complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the device was returned for analysis.Overall visual revision did not identify failures or evidence that could be lost due to decontamination process.Media inspection of a picture attached, and the reported error is not shown in the photo.Visual inspection of the accustick device was returned with the sheath, dilator and the canula stuck inside the device.It was observed that the ro marker was out of its location due it was separated from the accustick device.Additionally, the accustick device was damage at distal tip.No more damages were observed.The device was inspected under microscope, and it was observed that the distal tip was damage, this damage is located next to the ro marker position.
 
Event or Problem Description
It was reported that the markerband was dislodged.The target lesion was located in the kidney.An accustick ii w/o guidewire was selected for use.After withdrawing the device from the patient, it was noted that the silver markerband was found in the upper part of the dilator and not at the end of the introducer.The procedure was completed with this device.No complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUSTICK II
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18438597
Report Number2124215-2023-74573
Device Sequence Number20741341
Product Code KGZ
UDI-Device Identifier08714729157625
UDI-Public08714729157625
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/02/2026
Device Model Number35615
Device Catalogue Number35615
Device Lot Number0030775916
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/15/2023
Supplement Date Received by Manufacturer02/06/2024
Initial Report FDA Received Date01/03/2024
Supplement Report FDA Received Date02/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-