MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3544325 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section d 6b.Explantation date: (b)(6) 2024.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: material rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent breast prosthesis implantation with two mentor memorygel breast implants.Post-operatively, the patient was diagnosed, via mri, with bilateral breast implant leaks/ruptures.As a result, the patient has been scheduled for bilateral breast implant removal surgery on (b)(6) 2024.This medwatch form is for the left breast prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
On february 21, 2024, the mentor failure analysis lab received the device for evaluation.On february 27, 2024,, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the siltex hpg, 325cc breast implant had parallel lines of shell wear on the anterior view.In addition, a tear was noted on the posterior view extending to the anterior view within the shell wear measuring approximately 10.2 cm.The evaluation determined that the possible cause of the rupture was consistent with normal wear.Shell wear suggests in-vivo folding or creasing of the device.This may be the result of the following factors: continuous and sustained stresses to the device such as too small breast pocket and folding or wrinkling of the shell in the breast pocket.In some cases, the breast implants may also wear out over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.
|
| |
|
Additional Manufacturer Narrative
|
|
On january 22, 2024, mentor received additional information indicating that the patient was only diagnosed with a left breast implant rupture and that the right breast implant was identified to be intact upon removal.The patient's implants were replaced bilaterally with the following devices: (left) 300cc mentor memorygel breast implant catalog: 3503004bc, lot: 9961470 sn: (b)(6) and (right) 300cc mentor memorygel breast implant catalog: 3503004bc, lot: 9961470 sn: (b)(6).
|
| |
|