MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER
|
Back to Search Results |
|
| Model Number |
PED2-475-16 |
| Medical Device Problem Codes |
Improper or Incorrect Procedure or Method (2017); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received a report that the distal end of the pipeline would not open.Multiple attempts were made to get the device to open properly.The device was not deployed due to it not opening.The pipeline was placed in a distal bend.More than 50% of the pipeline had been deployed when it failed to open.The pipeline was resheathed more than 2 times.There were no additional steps or other devices required to open the pipeline.The pipeline was resheathed and removed from the patient with the microcatheter.The pipeline was used for an approved indication.The device and any accessories were prepared as indicated in the ifu.No patient symptoms or further complications were reported as a result of this event.The patient was undergoing surgery for treatment of a saccular, unruptured right internal carotid artery (ica) aneurysm.The landing zone was 3.3mm distally and 4.7mm proximally.It was noted the patient's vessel tortuosity was moderate.Angiographic result post procedure revealed successful placement of another pipeline device.Ancillary devices include a rist 079 ref 107f-079-95 ; lot 20101-01 sheath, phenom plus fg19120-1030-1s ; lot 225638850 guide catheter, phenom 27 fg15150-0615-1s ; lot 227525510 microcatheter.See manufacturer report 2029214-2024-00015 for related device.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis #800290618: equipment used: video inspection system (m-78210), ruler 200cm (m-83361).As found condition: the pipeline flex shield braid was returned for analysis within shipping box; within primary and secondary plastic bio-pouchs; within an opened pipeline e flex shield outer carton and inner pouches.The pushwire was not returned for analysis.Therefore any contributing factors from the pushwire could not be assessed.Damage location details: the braid was returned already detached from the pusher; therefore, the proximal and distal ends could not be identified.The distal and proximal ends of the pipeline flex shield were found fully opened and damaged.No other anomalies were observed.Testing/analysis: n/a.Conclusion: based on the analysis findings, the pipeline flex shield could not be confirmed to have failure/incomplete open distal (flex) as the device was resheated.In addition, the proximal and distal ends could not be identified.Damage to the braid identified during analysis, could have occurred due to resheathing of the pipeline flex shield more than two times.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|