| Catalog Number |
254500504 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : according to the information received, "[d254500504-attune insert tray gasket gets stuck and breaks during use during surgery]" the product was not returned to depuy synthes, however photos were provided for review.See attachment (b)(4).The photo investigation revealed that device 254500504, attune fb ps artic surf sz4 had broken surface and one of the post, deformed spring component, scratched and nicked surface.The failure mode is consistent with inserting an extractor device in between the trial and mating shim, and using the extractor to pry the mating devices apart during extraction of the trials.This improper technique results in damage to the spring and/or post components of the articulating surface as well as the mating shim.The attune intuition surgical technique 0612-10-512 (page 53), emphasizes the correct use of the tibial trial extractor (product code 254500138) with the trials.Furthermore, on page 51 of the surgical technique and per ifu-0902-00-836, careful inspection of the trials for damage/breakage should be performed pre and post-operative.If any damage to the balseal components is observed the trail should be replaced.In the event of instrument breakage during use, ensure that all device fragments that may have entered the surgical site are removed prior to completion of the procedure, as patient injury may result.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the device 254500504, attune fb ps artic surf sz4 would contribute to the complained device issue.Based on the investigation findings, potential cause can be attributed to unintended use error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot = > the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H6: component code: appropriate term/code not available (g07002) used to capture no findings available.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Attune insert tray gasket gets stuck and breaks during use in surgery.There was no patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthese joint reconstruction, or its employees caused or contributed to the potential event described in this report.(b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : according to the information received, "d254500504-attune insert tray gasket gets stuck and breaks during use during surgery".The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that attune fb ps artic surf sz4 has one of the post broken off, fragment was not returned for evaluation.Additionally, deep scratches were observed all over the surface, consistent with other tools and hard edges coming in contact with the device, properly handling and attention to the approved use of the device diminishes the risk of failure.The failure mode of the breakage is consistent with inserting an extractor device in between the trial and mating shim, and using the extractor to pry the mating devices apart during extraction of the trials.This improper technique results in damage to the spring and/or post components of the articulating surface as well as the mating shim.The attune intuition surgical technique 0612-10-512 (page 53), emphasizes the correct use of the tibial trial extractor (product code 254500138) with the trials.Furthermore, on page 51 of the surgical technique and per ifu-0902-00-836, careful inspection of the trials for damage/breakage should be performed pre and post-operative.If any damage to the balseal components is observed the trail should be replaced.In the event of instrument breakage during use, ensure that all device fragments that may have entered the surgical site are removed prior to completion of the procedure, as patient injury may result.The overall complaint was confirmed as the observed condition of the attune fb ps artic surf sz4 would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended user error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|
|