| Catalog Number |
SUPERHERO |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The account alleges that approximately 2-3 weeks post venous outflow component [voc] implant procedure, the voc implant had migrated out of the patient's right atrium and had rotated laterally inside the patient's arm just past the patient's shoulder.The physician elected to extract the voc along with the graft from the patient's arm.The patient is currently receiving dialysis via a central venous catheter.No patient injury to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device is not expected to return for evaluation.The complaint could not be confirmed.The root cause could not be determined.A review of the device history and complaint database could not be performed since the lot number was not provided.Should the device be returned later, the investigation will be reopened.
|
| |
|