MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 5F DL POWERMIDLINE MAX KIT; MIDLINE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 5F DL POWERMIDLINE MAX KIT; MIDLINE CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Codes Material Deformation (2976); Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation.H3 other text : device not returned.
 
Event or Problem Description
It was reported by the customer that picc kit had difficulty tearing and separating.Having to manipulate picc site leading to increased touches and risk for infection.Inability to place line is possible.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and is pending evaluation.Results are expected soon.
 
Event or Problem Description
It was reported by the customer that picc kit had difficulty tearing and separating.Having to manipulate picc site leading to increased touches and risk for infection.Inability to place line is possible.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of difficulty separating the microintroducer is confirmed; however, the root cause could not be determined.One photograph of a microintroducer was returned for evaluation.An initial visual observation of the photograph showed an implanted catheter and microintroducer.The t-handle was broken and the peel-apart sheath had been peeled most of the way.It appeared the t-handle did not completely break as a small piece of plastic was observed on the catheter.Although it appeared the microintroducer t-handle did not break evenly in the submitted photograph, inspection of the photograph was insufficient to identify the cause of the damage.Consequently, this complaint is confirmed as ¿cause unknown¿ at this time.This complaint will be recorded for future trending and monitoring purposes.H3 other text : evaluation findings are in section h11.
 
Event or Problem Description
It was reported by the customer that picc kit had difficulty tearing and separating.Having to manipulate picc site leading to increased touches and risk for infection.Inability to place line is possible.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5F DL POWERMIDLINE MAX KIT
Common Device Name
MIDLINE CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18441132
Report Number3006260740-2023-06006
Device Sequence Number6808544
Product Code PND
UDI-Device Identifier00801741129360
UDI-Public(01)00801741129360
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K162900
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model NumberN/A
Device Catalogue NumberP4255108D
Device Lot NumberREHW2072
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2023
Supplement Date Received by Manufacturer02/05/2024
03/18/2024
Initial Report FDA Received Date01/03/2024
Supplement Report FDA Received Date02/23/2024
03/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-