• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ADAPTA; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ADAPTA; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number ADDR01
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Heart Failure/Congestive Heart Failure (4446)
Date of Event 11/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient experienced persistent non-ischemic cardiomyopathy, which was presumed to be pacing induced, and a reduction in ejection fraction since implant of the implantable pulse generator (ipg).The patient was put on medical therapy and the ipg was explanted and replaced with a cardiac resynchronization therapy defibrillator (crt-d) system. no further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADAPTA
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18442218
Report Number3004209178-2024-00174
Device Sequence Number20521029
Product Code NVZ
UDI-Device Identifier00643169708624
UDI-Public00643169708624
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P980035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/14/2019
Device Model NumberADDR01
Device Catalogue NumberADDR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2023
Supplement Date Received by Manufacturer01/19/2024
Initial Report FDA Received Date01/04/2024
Supplement Report FDA Received Date02/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
457453, LEAD, 407458, LEAD.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age70 YR
Patient SexFemale
-
-