• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. KNOTLESS HIP FIBERTAK; NON-DEGRADABLE FIXATION FASTENER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. KNOTLESS HIP FIBERTAK; NON-DEGRADABLE FIXATION FASTENER Back to Search Results
Model Number KNOTLESS HIP FIBERTAK
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during a hip arthroscopy the suture tore off before the shuttling of the suture even started.Parts of the device remained inside the patient.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
The complaint allegation is confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The method of bone preparation and the quality of the bone encountered were not specified.The most likely cause for the reported failure can be attributed to user error of the device due to improper bone preparation, misalignment insertion, and/or prying/leveraging of the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KNOTLESS HIP FIBERTAK
Common Device Name
NON-DEGRADABLE FIXATION FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18444092
Report Number1220246-2024-00079
Device Sequence Number8145337
Product Code MBI
UDI-Device Identifier00888867295568
UDI-Public00888867295568
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K130458
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 01/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model NumberKNOTLESS HIP FIBERTAK
Device Catalogue NumberAR-3638H
Device Lot Number10560290
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2023
Supplement Date Received by Manufacturer12/27/2024
Initial Report FDA Received Date01/04/2024
Supplement Report FDA Received Date01/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-