| Model Number |
GF399R |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported to aesculap inc.On user facility medwatch report # (b)(4) that a suction tube tapered teardrop 12fr 180mm (part # gf399r) was used during an unspecified procedure on (b)(4).2023.According to the complainant, during the procedure, the device had broken.Reportedly, "all pieces were retrieved." the complaint device has not been returned to the manufacturer for evaluation.No patient complications were reported as a result of this event.The adverse event is filed under aic reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.Batch history review: the device history records (dhr) were not able to be reviewed as no lot number was made available.However, all device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.Conclusion and measures / preventative measures: a definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
|
| |
|
Search Alerts/Recalls
|