MAUDE Adverse Event Report: ZIMMER GMBH CMN FEMORAL NAIL, CCD 125°, LEFT, ø 10 MM, 21.5 CM; PROSTHESIS, TRAUMA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH CMN FEMORAL NAIL, CCD 125°, LEFT, ø 10 MM, 21.5 CM; PROSTHESIS, TRAUMA Back to Search Results
Model Number N/A
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/10/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).G2.Report source: france.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2024-00007.
 
Event or Problem Description
It was reported that a znn cmn nail was fitted on (b)(6) 2023 using the same ancillary reference whose cervical drill had broken a few weeks previously.During the cervical drill, it rubbed into the nail, leaving metal on the drill bit and in the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D10: znnâ¿¢, cmn lag screw reamer, short; item# 00249000364; lot# unknown.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.The lot number of the nail was reviewed for any deviations and/or anomalies in the manufacturing process that may have affected the surgical outcome or contributed to this reported event.No deviations or anomalies related to the reported event were discovered.Review of the complaint history for the nail found no additional related complaints for this item and the reported part and lot combination.An intraoperative fluoroscopy image was provided and reviewed by a radiologist.The image shows the left hip with an intramedullary nail already in place; the guiding pin is passed through the sleeve and through the nail hole, into the femoral head.Punctuate hyperdense foci adjacent to the guiding pin can be seen, most likely representing the metallic fragments described in the event description.Based on the available information, it is not possible to establish if the metallic debris left in the patient were produced by the nail, the lag screw reamer, the guiding pin or any combination of the three.In conclusion, a definitive root cause for the reported event could not be established.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that in a primary surgery a znn cmn nail was fitted using the same ancillary reference whose cervical drill had broken before.During the cervical drilling, it rubbed into the nail, leaving metal on the drill bit and in the patient.Diligence is complete and additional information on the reported event is unavailable at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CMN FEMORAL NAIL, CCD 125°, LEFT, ø 10 MM, 21.5 CM
Common Device Name
PROSTHESIS, TRAUMA
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18445856
Report Number0009613350-2024-00008
Device Sequence Number20520988
Product Code HSB
UDI-Device Identifier00889024298873
UDI-Public(01)00889024298873(17)330912(10)3175774
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K192312
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number47249321110
Device Lot Number3175774
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2023
Supplement Date Received by Manufacturer06/07/2024
Initial Report FDA Received Date01/04/2024
Supplement Report FDA Received Date06/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Other;
Patient SexFemale
Patient Weight47 KG
-
-