MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG ILIVIA NEO 7 DR-T DF-1 PROMRI; ICD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG ILIVIA NEO 7 DR-T DF-1 PROMRI; ICD Back to Search Results
Model Number 429529
Medical Device Problem Code Premature Elective Replacement Indicator (1483)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2023
Type of Reportable Event Malfunction
Event or Problem Description
This device shows eri with unexpected battery behavior.Device remains implanted.No adverse patient events were reported.Should additional information become available, this file will be updated.
 
Additional Manufacturer Narrative
(b)(6) 2024 device explanted.
 
Additional Manufacturer Narrative
On (b)(6) 2024 device explanted.Upon receipt, the device was interrogated, confirming the device status eri.The icd was implanted for 9 months and 7 charging cycles were recorded in the devices memory.The memory content of the device was inspected.The inspection revealed, that the icd activated the eri status, because during an automatic signature check the device detected invalid memory content in the live-holter, a memory range containing battery data.In general, the icd is equipped with the ability to detect and repair invalid memory content.If there was invalid memory content found in the live-holter, by design the icd will activate the device status eri to avoid an incorrect automatic longevity calculation.Furthermore, the ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.A fibrillation signal was applied and the device delivered a defibrillation shock as specified, documenting a normal and expected sensing and shock delivery.In particular, the specified energy level was reached and the charging time was normal and as expected.In conclusion, the clinical observation resulted from invalid memory content in the live-holter.The invalid memory content was automatically detected by the device leading to the activation of the eri status.The ability of the device to deliver therapies is not affected by this safety mechanism.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILIVIA NEO 7 DR-T DF-1 PROMRI
Common Device Name
ICD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key18446550
Report Number1028232-2024-00041
Device Sequence Number20748314
Product Code MRM
UDI-Device Identifier04035479156749
UDI-Public(01)04035479156749(17)241231
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model Number429529
Device Catalogue Number429529
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2024
Supplement Date Received by Manufacturer01/18/2024
10/09/2024
Initial Report FDA Received Date01/04/2024
Supplement Report FDA Received Date01/22/2024
10/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age70 YR
Patient SexMale
-
-