| Model Number |
M00576530 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/08/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Block a2: the patient's exact age is unknown; however, it was reported that the patient's year of birth is 1951.Block h6: imdrf device code a150101 captures the reportable event of stent failure to expand.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an ultraflex tracheobronchial covered distal stent was to be implanted in the left main stem of the lungs for the palliative treatment of lung cancer during a pulmonary tracheobronchial stent placement procedure performed on (b)(6) 2023.The patient's anatomy was not tortuous and was not dilated prior to stent placement.During the procedure, the stent was successfully deployed, but it did not expand correctly.The stent was removed from the patient with a pulmonary rat-tooth grasper, and another ultraflex tracheobronchial stent was used to complete the procedure.There were no patient complications reported as a result of this event.
|
| |
|