• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE FLOWPORT II CANNULA WITH OBTURATOR STRYKER 165MM; ACCESSORIES, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE FLOWPORT II CANNULA WITH OBTURATOR STRYKER 165MM; ACCESSORIES, ARTHROSCOPIC Back to Search Results
Catalog Number CAT02438
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the tip of the device broke during the procedure.The piece was retrieved.
 
Event or Problem Description
It was reported that the tip of the device broke during the procedure.The piece was retrieved.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: tip breaking.Probable root cause: design inadequate material selection to support movement/manipulation by user.Cannula tubing thickness too small to support movement/manipulation by user.Cannula size or geometry does not promote proper visualization of joint.Excessively sharp tip of cannula damages tissue.Manufacturing cannula not assembled, molded or machined to specification.Application excessive force poor visualization through arthroscope.The reported failure mode will be monitored for future reoccurrence.Manufacture date in not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOWPORT II CANNULA WITH OBTURATOR STRYKER 165MM
Common Device Name
ACCESSORIES, ARTHROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key18448367
Report Number0002936485-2024-00004
Device Sequence Number7660790
Product Code NBH
UDI-Device Identifier07613252633181
UDI-Public07613252633181
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCAT02438
Device Lot Number009484
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/08/2023
Supplement Date Received by Manufacturer12/08/2023
Initial Report FDA Received Date01/04/2024
Supplement Report FDA Received Date07/02/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-