| Model Number |
M00516740 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0501 captures the reportable event of tip detached.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a wallflex esophageal fully covered stent was to be implanted in the mid esophagus to treat a 2cm long malignant stricture during an esophagogastroduodenoscopy (egd) with esophageal stent placement procedure performed on (b)(6) 2023.During the procedure, the stent was advanced through the lesion; however, the tip got stuck in the tumor and got detached when the delivery system was attempted to be removed.The tip was removed from the patient using a rat tooth forceps and another wallflex esophageal stent was used to complete the procedure.There were no patient complications as a result of this event.
|
| |
|