| Catalog Number |
11-6400 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During post-operative radiographic imaging a pedicle screw yoke was observed to be un-engaged to the screw shank.A surgical procedure was performed on (b)(6) 2023 to remove and replace the hardware.
|
| |
|
Additional Manufacturer Narrative
|
|
The pedicle screw construct removed was returned for evaluation.Evaluation did not find any device issues related to this event.The exact cause could not be determined.Review of production and inspection records did not identify any issues relevant to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: correction to primary unique device identifier.
|
| |
|