TELEFLEX MEDICAL ET TUBE, SHER-I-BRONCH, LS, 37 FR; TUBE, TRACHEAL/BRONCHIAL, DIFF Back to Search Results
Catalog Number 5-16037
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that: on (b)(6) 2023, the doctor found cuff leakage when doing testing before using on patient.Then changed new one, no impact on patient.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
(b)(4).Customer complaint cannot be confirmed due the product sample is not available to perform a proper investigation and determinate the root cause.A device history record review was performed, and no relevant findings were identified.If the alleged defect samples become available at a later date, this complaint will be updated accordingly.
 
Event or Problem Description
It was reported that: (b)(6) 2023, the doctor found cuff leakage when doing testing before using on patient.Then changed new one, no impact on patient.
 
Additional Manufacturer Narrative
Qn#(b)(4).The reported complaint of "leak - balloon cuff" was confirmed based upon the sample received.The returned et tube was found to have a small hole in the bronchial (blue) balloon cuff which prevented it from being able to stay inflated.A device history record review was performed with no evidence to suggest a manufacturing related cause.All et tubes are 100% inspected for inflation and deflation at the time of manufacturing.It is unlikely that this type of damage was present at the time of manufacturing.Therefore, based upon the damage observed, unintentional user error caused or contributed to this event.
 
Event or Problem Description
It was reported that: (b)(6) 2023, the doctor found cuff leakage when doing testing before using on patient.Then changed new one, no impact on patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ET TUBE, SHER-I-BRONCH, LS, 37 FR
Common Device Name
TUBE, TRACHEAL/BRONCHIAL, DIFF
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
mariah mackinnon
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18451232
Report Number3003898360-2024-00062
Device Sequence Number7780778
Product Code CBI
UDI-Device Identifier04026704397617
UDI-Public04026704397617
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K180253
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number5-16037
Device Lot Number73K2200185
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2023
Supplement Date Received by Manufacturer01/15/2024
01/24/2024
Initial Report FDA Received Date01/05/2024
Supplement Report FDA Received Date01/17/2024
01/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-