• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS HCG+B; SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS HCG+B; SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN Back to Search Results
Catalog Number 03271749160
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The cobas e411 analyzer's serial number is (b)(6).The investigation is ongoing.
 
Event or Problem Description
There was an allegation of false positive elecsys hcg+b results for 1 patient sample on a cobas e411 module.The positive result was questioned due to it being impossible for the patient to be pregnant.The customer performed additional sample collections and still received positive results on the roche e411.Numeric result values were not provided.The customer tried processing the sample in a hydrophilic tube to remove anything that may cause a positive result, but the result was still positive.
 
Additional Manufacturer Narrative
The last calibration data that was provided was from (b)(6) 2023 and the results were slightly lower than expected.As the customer was unable to provide further information, a definite root cause analysis is not possible.Based on the data provided, the investigation could not identify a clear root cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS HCG+B
Common Device Name
SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18451285
Report Number1823260-2024-00042
Device Sequence Number18034082
Product Code DHA
UDI-Device Identifier04015630939022
UDI-Public04015630939022
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K003178
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Catalogue Number03271749160
Device Lot Number70917405
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/15/2023
Supplement Date Received by Manufacturer01/09/2024
Initial Report FDA Received Date01/05/2024
Supplement Report FDA Received Date01/16/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-