| Model Number |
PM2110 |
| Medical Device Problem Codes |
No Pacing (3268); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient had an auditory alert demonstrated and presented to the clinic.Upon review the pacemaker, it was no alert presented.No intervention was performed.The patient was stable and would be further monitored.
|
| |
|
Event or Problem Description
|
|
Additional information received notes that the pacemaker was explanted and replaced on (b)(6) 2024.
|
| |
|
Event or Problem Description
|
|
Additional information received notes that the pacemaker was explanted due to non-responsiveness.The patient remained in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported events of no pacing and noise audible could not be confirmed.The pacer was received in normal working conditions with battery voltage at elective replacement indicator (eri).Cautery marks were noted on pacer consistent with inhibited pacing during explant.Electrical and mechanical analysis performed, including output verification, indicated normal device functionality.When automatic settings are programed, the device output would adjust itself to accommodate the patient¿s needs for pacing and thus affect the estimated longevity of the device.The implant duration exceeds the projected longevity based on average monthly current indicating normal battery depletion.
|
| |
|