• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CERELINK ICP MONITOR; ICP MONITOR-CERELINK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CERELINK ICP MONITOR; ICP MONITOR-CERELINK Back to Search Results
Catalog Number 826820
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2023
Type of Reportable Event Malfunction
Event or Problem Description
A facility reported a cerelink monitor (id 314591) had negative values (-2 to -1) with appropriate troubleshooting.The negative values lasted about an hour at which time it reached around 2-3 where it had been before going negative.Product was in contact with the patient; however, no patient injury was reported.According to reporter: "the issue seems to be resolved so far based on the customer's feedback." no additional information has been provided after several attempts.
 
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The cerelink monitor (id 826820) was not returned for evaluation; therefore, an evaluation of the device could not be performed.Serial number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause(s) of the reported issue could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.Additional information: the customer mentioned that the patient lead was always attached.Unfortunately, they were unable to download the data since it has been used on a different patient since then.We have asked them to try to save the data if this issue arises again.
 
Additional Manufacturer Narrative
Follow-up report to include correction/removal rpt number (h9).
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERELINK ICP MONITOR
Common Device Name
ICP MONITOR-CERELINK
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
6099362319
MDR Report Key18452085
Report Number3014334038-2023-00199
Device Sequence Number20570393
Product Code GWM
UDI-Device Identifier10381780533788
UDI-Public10381780533788
Combination Product (Y/N)N
PMA/510(K) Number
K183406
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number826820
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/08/2023
Supplement Date Received by Manufacturer06/06/2024
08/08/2024
Initial Report FDA Received Date01/05/2024
Supplement Report FDA Received Date06/27/2024
08/09/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/03/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3014334038 -8/23/2022-00
Patient Sequence Number1
Patient SexUnknown
-
-