| Model Number |
6531 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6)2023.It was reported that shaft kink occurred.Upon unpacking of a 12mm x 3.00mm nc quantum apex balloon catheter, the shaft was noticed to be kinked.The procedure was completed with another device.No patient complications were reported.However, returned device analysis revealed hypotube detach.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of an nc quantum apex balloon catheter.The device was microscopically and visually examined.There was no sign of fluid within the device and the balloon was tightly folded.There were numerous kinks to the hypotube and shaft of the device.At 43.6cm from the strain relief, the hypotube of the device was separated.
|
| |
|