C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® I.C. ALL-SILICONE TEMP SENSING FOLEY CATHETER
|
Back to Search Results |
|
| Catalog Number |
119310M |
| Medical Device Problem Code |
Material Invagination (1336)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that customer having a few issues with bard temperature sensing catheters disconnecting from the uri meters recently.Customer asked they were just wondering if there have been any recent changes to the catheters as far as the plastics being used, any slight change in size of connection end, or anything else can think of that might be causing the issue.Per sample received on 02aug2023, customer returned 4 used all silicone foley catheters with material number 119310m.Per notification from investigator on 21dec2023, during sample evaluation it was found that the the balloon mushroom.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed cause unknown.One sample exhibited the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging), a 3-way temperature sensing catheter.Visual inspection noted balloon was not mushroomed unlike attached photos and there was a tear measuring 0.3235" found on the drainage arm.A photo was taken before evaluation was continued.Inflated balloon with 3.5 ml of solution and the catheter rested with no leaks noted.Deflated balloon passively and it was found that the balloon had mushroomed.This is out of specification, which states, " balloon must not cuff after deflation.A potential root cause for this failure mode could be balloon material does not shrink quickly enough.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Sterile: unless package is opened or damaged.Warning: do not use ointments or lubricants having a petrolatum base.They will damage silicone and may cause balloon to burst.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Single use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip syringe.Do not use needle.Recommended inflation capacities: 3cc balloon: use 5ml sterile water, 5cc balloon: use 10ml sterile water, 30cc balloon: use 35ml sterile water, do not exceed recommended capacities.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Note: aggressive traction, particularly in the presence of suturing, is not recommended for 100% silicone foley catheters.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|
Event or Problem Description
|
|
It was reported that customer having a few issues with bard temperature sensing catheters disconnecting from the uri meters recently.Customer asked they were just wondering if there have been any recent changes to the catheters as far as the plastics being used, any slight change in size of connection end, or anything else can think of that might be causing the issue.Per sample received on 02aug2023, customer returned 4 used all silicone foley catheters with material number 119310m.Per notification from investigator on 21dec2023, during sample evaluation it was found that the the balloon mushroom.
|
| |
|