|
|
| Catalog Number |
ECS33B |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190); Ejection Problem (4009)
|
| Health Effect - Clinical Codes |
Failure to Anastomose (1028); Unspecified Tissue Injury (4559)
|
| Date of Event |
12/08/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 1/5/2024.D4 batch # unk.H6: health effect ¿ clinical code gastrointestinal system (e10).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No lot or batch number was provided therefore a device history could not be done.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: what was the ¿malfunction¿ of the circular stapler? what were the indications for surgery? what surgical procedure was performed? did the patient receive any preoperative chemotherapy or radiation? were there any issues experienced with the device in the initial surgical procedure? what healthcare professional fired the device and what is his/her experience with the device? where in the green gap setting scale was the indicator located prior to firing (low-b, middle-b, or high-b)? did the healthcare professional wait 15 seconds after closing the device and then retighten prior to firing? was the device difficult to close? was the device difficult to fire? how may counter-clockwise revolutions of the adjusting knob were used to open the device? did the healthcare professional receive audible & tactile feedback when firing the device? were the donuts inspected? if so, please describe.Was a complete transection of the white breakaway washer visually confirmed? were there any issues noted with staple formation at any point throughout the care of the patient? if so, please describe the shape and location.What is the current status of the patient? does the surgeon believe the post-operative complications were related to an alleged deficiency of the device or were there other contributing factors to include patient tissue condition? was a leak test performed? if so, what type and what was the result? was the staple line visualized endoscopically during the initial surgical procedure? how many days postoperative did the leak occur? how was the leak identified? what was observed at the site of the leak upon reoperation? how was the leak addressed? were there any issues experienced with the device in the initial surgical procedure? does the surgeon believe the post-operative stenosis was related to an alleged deficiency of the device or were there other contributing patient factors? please explain.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a low anterior resection the end to end anastomosis broke down, possibly due to a malfunction of the circular stapler.Dr.Then had to perform an ileostomy.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 1/30/2024.Additional information was requested, and the following was obtained: what was the ¿malfunction¿ of the circular stapler? they could not get stapler out after firing.What were the indications for surgery? unknown.What surgical procedure was performed? lar.Did the patient receive any preoperative chemotherapy or radiation? unknown.Were there any issues experienced with the device in the initial surgical procedure? no.What healthcare professional fired the device and what is his/her experience with the device? (b)(6) assist has used before.Where in the green gap setting scale was the indicator located prior to firing (low-b, middle-b, or high-b)? unknown.They pulled circular out which tore the colon.The surgeon used sutures to complete.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 2/8/2024.D4: batch # 798a65.Investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the ecs33b device arrived with the anvil missing.The breakaway washer was not present and there were no staples present.The device was reloaded with staples, a new washer was placed on the device and it was tested for functionality with a test anvil.It fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: the firing stroke must be completed.Do not partially fire the instrument.Incomplete firing can result in malformed staples, incomplete cut line, bleeding, and leakage from the staple line and/or difficulty removing the device.Ensure that the firing trigger is fully squeezed to ensure proper staple formation and cutting of tissue.Open the device by turning the adjusting knob counter clockwise until the anvil shaft is fully exposed.Remove the anvil to expose the device trocar.Retract the device trocar by rotating the adjusting knob until the device trocar is no longer exposed.Look at the tissue compression scale and ensure the indicator is not in or near the green zone.The device may also be inserted without removing the anvil if the preferred application is a sutured purse-string technique.In this case, however, prior to insertion, ensure the anvil is properly attached and the device is fully closed by rotating the adjusting knob clockwise.For removal.To safely release the device from the newly formed anastomosis, turn the adjusting knob counter-clockwise for two complete revolutions.The event described could not be confirmed as the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch number 798a65, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|