• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER GMBH BLUNT TIP SCREW, ÿ 4X38MM; PROSTHESIS, TRAUMA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH BLUNT TIP SCREW, ÿ 4X38MM; PROSTHESIS, TRAUMA Back to Search Results
Model Number N/A
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/07/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that operation was performed with ann approximately 1 month ago.After 4 days from the initial, surgeon found all proximal screws were backed out from the proper position.No reoperation date has been set, but the ann will be removed, and the rsa will be performed.No known impact or consequence to the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).D10.Cortical bone screw, 4x28mm item# 47248612840 lot# 3152732.Cortical bone screw, 4x28mm item# 47248612840 lot# 3164731.Proximal humerus nail cap, 0mm item# 47248801000 lot# 3178851.G2.Report source: japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2024-00010, 0009613350-2024-00011, 0009613350-2024-00013.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.A review of the device manufacturing records confirmed no abnormalities or deviations.Devices are used for treatment.Review of complaint histories identified additional similar complaints for the reported products and additional similar complaints for the reported part and lot combinations.Complaints are monitored through monthly complaint review, in order to identify potential adverse trends.X-rays received: images assessed, not sending to mmi for unable to coordinate undated images to event, sending images would not enhance the investigation.No surgical report was provided.It can be assumed that multiple contributing factors, related to either patient condition and behavior or implantation procedure, contributed to the migration of the screws.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.An additional deeper investigation was performed which identified the design limitation of the corelock mechanism of the affixus nail as a potential contributing factor.As part of the deeper investigation, a scientific literature search was conducted and a systematic review of the outcome of intramedullary nailing for acute proximal humerus fracture showed that screw migration and perforation into the joint is the second most common complication following secondary loss of reduction.In addition, scientific literature of competitor and predicate devices were assessed.This review has shown that the affixus® natural nail® proximal humeral system performs better and/or in equal range in comparison with legally marketed similar devices.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2024-00010-1.0009613350-2024-00011-1.0009613350-2024-00013-1.
 
Event or Problem Description
Diligence is complete and additional information on the reported event is unavailable at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLUNT TIP SCREW, ÿ 4X38MM
Common Device Name
PROSTHESIS, TRAUMA
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18455903
Report Number0009613350-2024-00012
Device Sequence Number7066532
Product Code HSB
UDI-Device Identifier00889024505438
UDI-Public(01)00889024505438(17)280816(10)3169752
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181827 / K2
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number47248603840
Device Lot Number3169752
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2023
Supplement Date Received by Manufacturer02/05/2024
Initial Report FDA Received Date01/05/2024
Supplement Report FDA Received Date02/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexFemale
-
-