| Model Number |
PCO2520FX |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Codes |
Hernia (2240); Tissue Breakdown (2681); Unintended Radiation Exposure (4565)
|
| Date of Event |
12/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but two photos were available for evaluation.The two images showed a recurrent hernia.The mesh was well placed into the abdominal wall.The textile knitting pattern matched with a parietex optimized composite mesh.A hole was visible on the mesh, where the hernia recurrence was located.It was reported that the mesh was torn.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the mesh should be used in the form in which it is provided without being cut.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, post operatively, there were no problems during the surgery in 2014.This time, the patient was examined for a recurrence of ventral hernia, and the x-rays showed that the hernia had been formed from the center of the mesh, however, during a repeat surgery, the center of the mesh was found to have been damaged and it was found that the hernia was starting from that hole.A mesh placement with intraperitoneal onlay mesh repair (ipom) plus.
|
| |
|
Search Alerts/Recalls
|
|
|