• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LDR MEDICAL ROI-C COATED IMPLANT 14X15.5 H6; ROI-C TITANIUM-COATED IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LDR MEDICAL ROI-C COATED IMPLANT 14X15.5 H6; ROI-C TITANIUM-COATED IMPLANT SYSTEM Back to Search Results
Catalog Number CMC1322P
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1 phone number: (b)(6).Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event or Problem Description
It was reported that the second plate of an roi-c construct could not be installed within the spacer intra-operatively.During installation of the second construct at the adjacent level, a piece of the anterior portion of the spacer broke off and was wedged in the channel between the spacer and the plate; the piece could not be recovered.There was a delay of 30 minutes to the procedure, but there have been no reported patient impacts.This is report two of two for this event.
 
Additional Manufacturer Narrative
If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference report (b)(4).
 
Event or Problem Description
It was reported that the second plate of an roi-c construct could not be installed within the spacer intra-operatively.During installation of the second construct at the adjacent level, a piece of the anterior portion of the spacer broke off and was wedged in the channel between the spacer and the plate; the piece could not be recovered.There was a delay of 30 minutes to the procedure, but there have been no reported patient impacts.This is report two of two for this event.
 
Event or Problem Description
It was reported that the second plate of an roi-c construct could not be installed within the spacer intra-operatively.During installation of the second construct at the adjacent level, a piece of the anterior portion of the spacer broke off and was wedged in the channel between the spacer and the plate; the piece could not be recovered.There was a delay of 30 minutes to the procedure, but there have been no reported patient impacts.This is report two of two for this event.
 
Additional Manufacturer Narrative
H6: additional method code: 4109.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the product was not returned so an evaluation was unable to be performed, however photos were provided.The provided photos show that a inferior plate was not installed but do not clearly show that the cage was fractured so the complaint remains unrefuted.Root cause: a definitive root cause cannot be determined with the information provided.The second anchoring plate was unable to advance through the cage because the cage had fractured and a piece was jammed in the anchoring plate pathway.The cage fracture could be related to difficulty implanting the first anchoring plate.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROI-C COATED IMPLANT 14X15.5 H6
Common Device Name
ROI-C TITANIUM-COATED IMPLANT SYSTEM
Manufacturer (Section D)
LDR MEDICAL
5 rue de berlin
quartier europe de l'ouest
sainte-savine 10300
FR  10300
Manufacturer (Section G)
LDR MEDICAL
5 rue de berlin
quartier europe de l'ouest
sainte-savine 10300
FR   10300
Manufacturer Contact
sabrina abla
10225 westmoor dr.
westminster, CO 80021
7206965158
MDR Report Key18456320
Report Number3004788213-2024-00145
Device Sequence Number16922911
Product Code OVE
UDI-Device Identifier03662663023184
UDI-Public(01)03662663023184(17)270401(10)C91032
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151934
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCMC1322P
Device Lot NumberC91032
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date01/05/2024
Supplement Report FDA Received Date01/26/2024
03/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
Patient EthnicityNon Hispanic
-
-