Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.
|
|
The user facility reported to terumo cardiovascular that prior cardiopulmonary bypass, during other - non clinical activity, upon unpacking and inspection, particulate in the material (housing) was detected.No patient involvement.
|
|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on january 5, 2024.Upon further investigation of the reported event, the following information is new and/or changed: d9 (device availability - added date returned to manufacturer) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information and device evaluation) h3 (device evaluated by manufacturer) h4 (device manufacture date) h6 (identification of evaluation codes 10, 11, 3331, 213, 67) the affected sample was inspected upon receipt to confirm a black mark unit in the reservoir.Further inspection revealed that the marking was embedded within the plastic.The mark was measured and found to be 0.2mm² which is within specification.A representative retention sample was inspected to show no anomalies with the device, including no foreign materials.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|