| Catalog Number |
MX448HFB |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D4: lot number, expiration date and h4: device manufacture date is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that upon visual inspection during first care time, yellow fluid noted on the sheets underneath the tri-fuse area.Provider notified about leaking tubing/connection.Pads placed under lines, fluid apparently leaking from connection between tpn filter and anti-siphon valve.Tpn/il taken down and new clear fluid hung.There was patient involvement and unknown patient harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Supplemental generated to correct previous report 9616567-2024-00002-001 due date.Aware date should have been 1/18/2024 with due date of 2/17/2024.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for investigation.No damage or anomalies were observed with the device during visual inspection.The device was functionally tested, and no leakage was observed.No product lot number was provided, therefore, no review of manufacturing device history records could be conducted.Based on the results of the functional testing, the investigation could not confirm the reported issue.As a preventative measure, manufacturing personnel were notified of this complaint.
|
| |
|