MAUDE Adverse Event Report: SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. NON-DEHP EXT SET 1.2 MICRON FILTER APV=1.0ML 50/CA; FILTER, INFUSION LINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. NON-DEHP EXT SET 1.2 MICRON FILTER APV=1.0ML 50/CA; FILTER, INFUSION LINE Back to Search Results
Catalog Number MX448HFB
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/29/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D4: lot number, expiration date and h4: device manufacture date is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that upon visual inspection during first care time, yellow fluid noted on the sheets underneath the tri-fuse area.Provider notified about leaking tubing/connection.Pads placed under lines, fluid apparently leaking from connection between tpn filter and anti-siphon valve.Tpn/il taken down and new clear fluid hung.There was patient involvement and unknown patient harm/adverse event reported.
 
Additional Manufacturer Narrative
Supplemental generated to correct previous report 9616567-2024-00002-001 due date.Aware date should have been 1/18/2024 with due date of 2/17/2024.
 
Additional Manufacturer Narrative
One device was received for investigation.No damage or anomalies were observed with the device during visual inspection.The device was functionally tested, and no leakage was observed.No product lot number was provided, therefore, no review of manufacturing device history records could be conducted.Based on the results of the functional testing, the investigation could not confirm the reported issue.As a preventative measure, manufacturing personnel were notified of this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NON-DEHP EXT SET 1.2 MICRON FILTER APV=1.0ML 50/CA
Common Device Name
FILTER, INFUSION LINE
Manufacturer (Section D)
SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
carretera miguel alemán km21.7
apodaca nuevo leon
MX 
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18457460
Report Number9616567-2024-00002
Device Sequence Number7079419
Product Code FPB
UDI-Device Identifier10351688507457
UDI-Public10351688507457
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K860803
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMX448HFB
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Returned to Manufacturer01/05/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2023
Supplement Date Received by Manufacturer12/06/2023
01/18/2024
Initial Report FDA Received Date01/05/2024
Supplement Report FDA Received Date02/14/2024
02/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-