| Catalog Number |
1400450-12 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulty appears to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a 90% stenosed lesion in the right coronary artery (rca) with heavy calcification and moderate tortuosity.The 4.5x12mm nc trek neo balloon dilatation catheter (bdc) was advanced to the lesion without resistance; however, the balloon ruptured at 18 atmospheres during the second inflation.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
|
| |
|