| Model Number |
71005 |
| Medical Device Problem Codes |
Fracture (1260); Migration (4003)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
12/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The patient reported cessation of therapy occurred after sustaining multiple falls.The system had been in use for more than a year with no apparent issues prior to the falls.It is likely that the external trauma directly affected the implanted system and led to the fracture and migration of the leads.
|
| |
|
Event or Problem Description
|
|
Patient was implanted with the nalu spinal cord stimulator system on (b)(6) 2022.Patient had reported receiving good therapy from the system after implant.In (b)(6) 2023 the patient reported having sustained multiple falls and subsequent loss of pain relief from the nalu system.Imaging and impedance checks determined that one of the implanted leads had fractured and the other implanted lead had migrated away from the target area for pain relief.On (b)(6) 2023 a surgical revision was performed in which all of the implanted components were replaced and new components were implanted at the target location.
|
| |
|