| Model Number |
CLK-4 |
| Medical Device Problem Code |
No Visual Prompts/Feedback (4021)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation and the evaluation found unit powered but failed to turn on the lamp due to faulty thermal fuse and main lamp not working.Spare lamp was found to be missing.Additionally, front panel was damage with crack at the bottom, bottom chassis was rusted, air pressure low due to faulty air pump unit.The investigation is ongoing and follow-up with user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus, the halogen light source had lamp lighting issues.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation it was confirmed the issue of ¿lamp not lighting¿ was due to a faulty thermal fuse.However, the specific cause of the faulty thermal fuse was not established at this time.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|