Model Number 0638-803-01
Medical Device Problem Code Output below Specifications (3004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no device available for analysis; therefore, no physical or visual analysis of the product could be performed.Based on the information available, a conclusion code cause not established was assigned to this investigation.
 
Event or Problem Description
It was reported that during the procedure, the console displayed case saver mode and entered in lower power setting, so the procedure was cancelled.The patient was stable under general anesthesia, there was no patient complication.After that, the company representative informed that the machine was usable for very small stones at low power, so we were able to complete the next case without any issues.This event is being reported for aborted/cancelled procedure with a patient under anesthesia or sedation is unknown.
 
Additional Manufacturer Narrative
The console or components were not returned for analysis.However, the console was evaluated by a field service engineer (fse) at the customer site.During functional evaluation of the console, there was found leakage residue at brick 2 and 3.After the field service engineer replaced the bricks, the issue was resolved, and the unit was within specification.As a result, the console was functioning as intended.Most likely the damaged bricks could have affected the device performance, leading to the reported event.Therefore, the reported allegation was confirmed.Based on review of all information available and analysis results, a conclusion code of cause traced to component failure was assigned to this investigation.
 
Event or Problem Description
It was reported that during the procedure, the console displayed case saver mode and entered in lower power setting, so the procedure was cancelled.The patient was stable under general anesthesia, there was no patient complication.After that, the company representative informed that the machine was usable for very small stones at low power, so we were able to complete the next case without any issues.This event is being reported for aborted/cancelled procedure with a patient under anesthesia or sedation is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPULSE POWERSUITE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
LUMENIS LTD
hakidma 6
yokneam 20692 04
IS  2069204
Manufacturer (Section G)
LUMENIS LTD
yokneam industrial park
hakidma st 6
yokneam 20692 04
IS   2069204
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18461613
Report Number2124215-2023-74057
Device Sequence Number7066289
Product Code GEX
UDI-Device Identifier07290109146195
UDI-Public07290109146195
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0638-803-01
Device Catalogue Number0638-803-01
Device Lot Number0000000536
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/18/2023
Supplement Date Received by Manufacturer01/24/2024
Initial Report FDA Received Date01/07/2024
Supplement Report FDA Received Date02/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/21/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-