| Catalog Number |
0700615 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiration date: 03/2028) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
| |
|
Event or Problem Description
|
|
It was reported that one month and twenty-one days post a chronic catheter placement, the red lumen allegedly broke.It was further reported that fluid allegedly leaked right where the lumen meets the red portion of the end of the line.Reportedly, the catheter was removed and replaced.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one powerline 6f d/l catheter was returned for evaluation.Functional, gross visual, tactile and microscopic visual evaluations were performed.The complete circumferential break on the distal end of the catheter were noted to be uneven and also the surface was noted to be granular.No break or leaks were noted on red and purple color luer.Therefore, the investigation is confirmed for the reported fracture as circumferential break was noted on the distal end of the catheter, however no anomalies were noted to both red and purple luers.Also, the investigation is unconfirmed for the reported fluid leak issue as no objective evidence found during sample evaluation.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 03/2028), g3 h11: h6 (method, result, conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported that one month and twenty-one days post a chronic catheter placement, the red lumen allegedly broke.It was further reported that fluid allegedly leaked right where the lumen meets the red portion of the end of the line.Reportedly, the catheter was removed and replaced.There was no reported patient injury.
|
| |
|