Model Number MI1000 MED-EL CONCERT
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2023
Type of Reportable Event Malfunction
Event or Problem Description
The user visited the clinic with bloody ear discharge and no more hearing benefit.Upon examination, the doctor observed that the ear was wet and noticed the extrusion of the electrode lead, which was visible in the ear canal.As a result, the doctor recommended re-implantation.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The user visited the clinic with bloody ear discharge and no more hearing benefit with the device.The problem started 10-12 days prior (mid december 2023).The user has been reimplanted with a new device.
 
Additional Manufacturer Narrative
Conclusions: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.According to the information received device was explanted due to an extrusion of the electrode lead into the external ear canal.Reportedly the main reason for the extrusion was due to a device unrelated enlarged cholesteatoma and granulation.In addition, the recipient suffered from an ear infection.A review of the device_s sterilization records shows that the device has been subject to a valid sterilization process, thus no available information points to the implant being the source of infection.The user has history of ear infections prior to implantation.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key18462578
Report Number9710014-2024-00012
Device Sequence Number15240156
Product Code MCM
UDI-Device Identifier09008737062682
UDI-Public(01)09008737062682
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/10/2020
Device Model NumberMI1000 MED-EL CONCERT
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2023
Supplement Date Received by Manufacturer12/28/2023
Initial Report FDA Received Date01/08/2024
Supplement Report FDA Received Date03/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age9 YR
Patient SexFemale
-
-