• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB STELLARIS ELITE ADAPTIVE FLUIDICS STANDARD PACK; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB STELLARIS ELITE ADAPTIVE FLUIDICS STANDARD PACK; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number BL5115-2
Medical Device Problem Code Particulates (1451)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned and a lot number was not provided; therefore, the sterilization and lot history records could not be reviewed.The lot number has been requested.The investigation is ongoing.
 
Event or Problem Description
The user facility in germany reported that the site-hole of the sleeve was not entirely stamped out and a little yellow particle fell off the sleeve and entered the patient ''s eye.The particle was removed using a cannula aspiration and removed after incision.The surgery was finished successfully with no impact to the patient.This report is for the first patient from the same facility.The surgery was finished successfully.No patient impact.
 
Additional Manufacturer Narrative
The product nor the particle was available for evaluation and no further information has been provided.Therefore, we are unable to investigate further for root cause.The trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLARIS ELITE ADAPTIVE FLUIDICS STANDARD PACK
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
BAUSCH + LOMB
1400 n goodman st
rochester NY 14609
Manufacturer (Section G)
BAUSCH & LOMB INCORPORATED
3365 tree ct. industrial blvd
saint louis MO 63122
Manufacturer Contact
juli moore
3365 tree court industrial blvd
st. louis, MO 63122
6362263220
MDR Report Key18462760
Report Number0001920664-2023-70152
Device Sequence Number7079692
Product Code HQC
UDI-Device Identifier00757770058853
UDI-Public(01)00757770058853
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K170052
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBL5115-2
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/08/2024
Supplement Report FDA Received Date01/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-