MAUDE Adverse Event Report: CORDIS US CORP. TEMPO; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS US CORP. TEMPO; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 451513H0
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
As reported, a 5f tempo berenstein (ber) 100cm diagnostic catheter ripped at its midpoint from the tip after insertion into the left femoral artery.The catheter was safely removed from the patient¿s body.This information is sourced from medwatch mw5147733.Additional event details are not available.
 
Additional Manufacturer Narrative
As reported, a 5f tempo berenstein (ber) 100cm diagnostic catheter ripped at its midpoint from the tip after insertion into the left femoral artery.The device was not returned for evaluation.Without the return of the device or images for analysis, the reported customer event ¿catheter (body/shaft)-cracked¿ could not be confirmed.Shipping/handling factors may have contributed to the reported event.Users are trained to inspect for signs of damage prior to and during use.Any product with damage is not to be used.Information for safety is provided in the products labeling with the intent to make the user aware of the risks.According to the instructions for use (ifu), although not intended as a mitigation of risk, ¿to prevent damage to the catheter tip during removal from the package, grasp the hub and withdraw the catheter.Treat all 4f (1.35 mm) catheters and smaller french sizes with ultimate care.The performance of these products may be impaired if not properly and cautiously handled during unpacking and preparation.¿ based on the available information, there is no indication that the event is related to the device design or manufacturing process.Therefore, no preventive or corrective actions will be taken at this time.
 
Additional Manufacturer Narrative
After further review, section d.4 primary unique device identification (udi) number has been corrected accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPO
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS US CORP.
14201 nw 60th avenue
miami lakes, florida 33014
Manufacturer (Section G)
CORDIS US CORP.
14201 nw 60th avenue
miami lakes, florida 33014
Manufacturer Contact
karla castro
santiago troncoso 808
juarez, chihuahua 
7863138372
MDR Report Key18465584
Report Number9616099-2024-00009
Device Sequence Number7065934
Product Code DQO
UDI-Device Identifier10705032008440
UDI-Public(01)10705032008440(17)260331(10)18199504
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K973401
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Catalogue Number451513H0
Device Lot Number18199504
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2023
Supplement Date Received by Manufacturer08/15/2024
Initial Report FDA Received Date01/08/2024
Supplement Report FDA Received Date08/15/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.
Patient SexUnknown
-
-