| Model Number |
CLK-4 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A user facility returned the olympus asset, the halogen light source, to the olympus service center.Upon inspection and testing of the returned device, it was observed that the unit is not turning on due to misalignment of the top cover.This report is being submitted for the reportable malfunction found during evaluation.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned as part of the asset return process which found the cover was misaligned causing the unit not to turn on.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation the specific cause of the reported issue of ¿device cannot be turned on¿ could not be identified.However, misalignment of the top cover may have contributed to the reported issue.Olympus will continue to monitor field performance for this device.
|
| |
|