• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSTEOMED, LLC 2.0MM X 6MM AUTODRIVE SCREW; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSTEOMED, LLC 2.0MM X 6MM AUTODRIVE SCREW; SCREW, FIXATION, BONE Back to Search Results
Model Number 211-2006
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices were received on 19 december 2023 for evaluation.Review of the dhrs revealed no anomalies.Additionally, review of material composition of the devices as well as manufacturing processes indicated no changes to either since release of these products.As received, the four (4) plates were fractured, and visual inspection revealed some warping on the plates.It is unknown if this warping occurred due to the reported event, during the initial implant surgery, or due to explant damage.Visual inspection of the returned screws revealed no anomalies.The returned parts were given to qc inspection for evaluation against the engineering drawings, and all devices met specifications.A two-year review of the complaint database for the 211-20xx part family revealed only one (1) complaint for post-op plate breakage- which is the complaint within this report.Similarly, review of the complaint database for part number 210-0030 revealed one (1) complaint for post-op plate breakage (which is the complaint in this report).Potential causes for post-op plate breakage include improper material selection/strength, user error-improper fixation technique, patient noncompliance with post-op instructions, excessive bending while contouring plates, improper plate size selection, and no/slow osteogenesis, osteolysis, bone resorption.However, based on the information received and the investigation performed, the root cause could not be determined.Original report was submitted: report number 2027754-2023-00054.Based on the additional information received on 22 december 2023, it was reported additional devices were involved, therefore, the following related reports have been submitted: 2027754-2024-00002, 2027754-2024-00003, 2027754-2024-00004, 2027754-2024-00005, 2027754-2024-00006, 2027754-2024-00007, 2027754-2024-00008, 2027754-2024-00009, 2027754-2024-00010, 2027754-2024-00017, 2027754-2024-00018, 2027754-2024-00019, and 2027754-2024-00020.
 
Event or Problem Description
Additional information was received on 22 december 2023 regarding the patient and the reported event.It was reported the original implant surgery was on (b)(6) 2023, and 12 days after surgery (event date (b)(6) 2023), a radiographic examination was performed where it was confirmed that the fixed plates were broken.On (b)(6) 2023, a new surgery was performed, where the material was explanted, and a new material was implanted again.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM X 6MM AUTODRIVE SCREW
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer (Section G)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer Contact
ellie wood
3885 arapaho rd
addison, TX 75001
9726774600
MDR Report Key18466411
Report Number2027754-2024-00016
Device Sequence Number3087380
Product Code HWC
UDI-Device Identifier00845694002448
UDI-Public(01)00845694002448(10)1172397(30)1(11)220802
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number211-2006
Device Catalogue Number211-2006
Device Lot Number1172397
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date01/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-