• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 HP UNIVERSAL HANDLE; KNEE INSTRUMENT : HANDLES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 HP UNIVERSAL HANDLE; KNEE INSTRUMENT : HANDLES Back to Search Results
Catalog Number 950501305
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
This instrument is composed of 1 part.The knob of the instrument has snapped off.The product damage documented in this pc has been identified during loan kit inspection, by the loan kit technician.  the event date  and  alert date are the date that the inspection took place.  there is no surgeon, procedure, or patient details available. .
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: this instrument is composed of 1 part.The knob of the instrument has snapped off.As per the images attached.The product damage documented in this pc has been identified during loan kit inspection, by the loan kit technician.The event date and alert date are the date that the inspection took place.There is no surgeon, procedure, or patient details available.No further information can be obtained as the case was not reported by the customer.The device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the returned device found that the hp universal handle was broken from the cylindrical tip.The broken fragment was returned for evaluation.The reported condition can be confirmed.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the hp universal handle would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code (a1107) provided is not a valid finished goods lot#.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: this instrument is composed of 1 part.The knob of the instrument has snapped off.As per the images attached.The product was not returned to depuy synthes, however photos were provided for review.See attachment (950501305.Jpg).The photo investigation revealed that hp universal handle distal cylindrical tip had fractured at the interface with the square cross section.The reported condition can be confirmed.No other issues were found.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the hp universal handle would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a date code was provided, which indicates that the device was manufactured on (11/15/2007).A manufacturing records evaluation (mre) was not performed since a valid finished good lot number was not provided for this device.
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: this instrument is composed of 1 part.The knob of the instrument has snapped off.The product was not returned to depuy synthes.However, photos were provided for review.The photo investigation revealed, that hp universal handle distal cylindrical tip had fractured at the interface, with the square cross section.The reported condition can be confirmed.No other issues were found.The observed, condition was identified, as an end of life indicator.Damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use.The device must be properly inspected, prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions and inspection procedures.As the device was not returned.An as-received condition could not be assessed.And a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed.As the observed, condition of the hp universal handle would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life.And it has been determined, that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a date code was provided.Which indicates, that the device was manufactured on (11/15/2007).A manufacturing records evaluation (mre) was not performed, since a valid finished good lot number was not provided for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HP UNIVERSAL HANDLE
Common Device Name
KNEE INSTRUMENT : HANDLES
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18466787
Report Number1818910-2024-00578
Device Sequence Number19604923
Product Code LXH
UDI-Device Identifier10603295226031
UDI-Public10603295226031
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number950501305
Device Lot NumberA1107
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer01/16/2024
03/06/2024
03/21/2024
03/25/2024
Initial Report FDA Received Date01/08/2024
Supplement Report FDA Received Date01/17/2024
03/07/2024
03/22/2024
03/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/15/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-