| Catalog Number |
09180443001 |
| Medical Device Problem Codes |
Application Program Problem (2880); Data Problem (3196)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of a software issue for the cobas infinity lab core license 3.X.The reporter indicated the system was not displaying the unit of measure for the previous result on the validation screen.The unit of measure is only visible for the current value.When connected to the cobas 8800 module, quantitative results with 3 or more digits are expressed in the form "3 digits + 10*x iu/ml".The software version is 3.05.No patient was harmed due to the issue.The results were not reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
Section e1 correction: the initial reporter's first and last name have been translated to english.H6 medical device problem, investigation findings, and investigation conclusion codes were updated.During the investigation, the allegation was reproduced on a blank environment using a cobas 8800.It was observed that the units of the 'last repeated result' were not displayed in cobas infinity causing comprehension issues in case the units are different from the actual one.The reported allegation has been verified as a software issue in cobas infinity, as the units of the 'last repeated result' value are not displayed in infinity causing comprehension issues in case the units are different from the actual one.
|
| |
|