• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERRAGENE S.A STRIKER; BIOLOGICAL INDICATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERRAGENE S.A STRIKER; BIOLOGICAL INDICATOR Back to Search Results
Model Number STRYKER SRBI30
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there were recurring bis failing following completed vp4 cycles.The vp4 did not register any error codes or cancel any loads.
 
Additional Manufacturer Narrative
No device-related deaths or serious injuries were informed.Device malfunction cause the device to fail to perform its essential function and compromise the device's monitoring effectiveness, since it would not be able to process biological indicators.If this were to recur, the lack of biological indicators processing could affect the load release, which could contribute to a serious injury due to lack of sterile material if it were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRIKER
Common Device Name
BIOLOGICAL INDICATOR
Manufacturer (Section D)
TERRAGENE S.A
ruta nacional no. 9, km 280
parque industrial micropi
alvear, santa fe 2130
AR  2130
Manufacturer (Section G)
TERRAGENE S.A
ruta nacional no. 9, km 280
parque industrial micropi
alvear, santa fe 2130
AR   2130
Manufacturer Contact
hernando carrizo
ruta nacional no. 9, km 280
parque industrial micropi
alvear, santa fe 2130
AR   2130
MDR Report Key18467518
Report Number3013145340-2024-00013
Device Sequence Number7192399
Product Code FRC
UDI-Device Identifier07798164678793
UDI-Public07798164678793
Combination Product (Y/N)N
PMA/510(K) Number
K191021
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberSTRYKER SRBI30
Device Lot NumberUNKNOWN
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2022
Initial Report FDA Received Date01/08/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-