| Model Number |
STRYKER SRBI30 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there were recurring bis failing following completed vp4 cycles.The vp4 did not register any error codes or cancel any loads.
|
| |
|
Additional Manufacturer Narrative
|
|
No device-related deaths or serious injuries were informed.Device malfunction cause the device to fail to perform its essential function and compromise the device's monitoring effectiveness, since it would not be able to process biological indicators.If this were to recur, the lack of biological indicators processing could affect the load release, which could contribute to a serious injury due to lack of sterile material if it were to recur.
|
| |
|
Search Alerts/Recalls
|