| Model Number |
3006 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation, and the additional findings were as follows: dried tissues were observed adhered to both the distal tip and blade, the lock mechanism was found in the "off" position and there was noticeable bending at the distal end, and there were multiple cracks noted in the flare.The investigation is ongoing, and follow up with the facility is being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus, that the cutting forceps had a potential out of box failure upon initial use (there was no ntome for the seal).The device was returned for evaluation.During the device evaluation, the jaws were unresponsive to manipulation of the handle and remained closed without opening.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: e2 and e3 were inadvertently omitted from the initial report.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The jaws were unresponsive to manipulation of the handle, remaining closed without opening.Additionally, multiple cracks were noted in the flare.Unfortunately, a functional test was unable to be performed due to the non-opening nature of the jaws.The event was unable to be confirmed as the device was damaged when it was received by olympus.The root cause of the damage was unable to be determined.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|