Catalog Number 254500930
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the silver ring popped out from attune shim.Patient status/ outcome / consequences ? , was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: ? unknown, is the patient part of a clinical study? unknown, (b)(4).Device property of? none, device in possession of? none.By checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst? true.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : according to the information received, "the silver ring popping out from attune shim.As per the images attached".The product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that attune conv shim sz5 14mm metal rings are slightly out of their place.Additionally a thin line around both metal rings that seems to be a crack was observed.It is possible that the condition of the rings was caused by the cracks.Due to crack propagation being recognized as an inherent risk of the shims after repeated use, the potential cause is traced to end of life.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the attune conv shim sz5 14mm would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, "the silver ring popping out from attune shim.As per the images attached".The product was not returned to depuy synthes, however photos were provided for review.See attachments 254500930.Jpg and 254500930 2nd.Jpg.The photo investigation revealed that attune conv shim sz5 14mm metal rings are slightly out of their place.Additionally a thin line around both metal rings that seems to be a crack was observed.It is possible that the condition of the rings was caused by the cracks.Due to crack propagation being recognized as an inherent risk of the shims after repeated use, the potential cause is traced to end of life.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the attune conv shim sz5 14mm would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary - according to the information received, the silver ring popping out from attune shim.The device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the returned sample revealed that attune conv shim sz5 14mm shows that both rings are slightly protruding out of their place.Additionally, a thin line around both metal rings that seems to be a crack was observed.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The thermal properties of the metal bushing and radel trial body are different resulting in the material to expand and contract at different rates.This puts a significant amount of stress on interfacing materials.Radel, being highly brittle, will succumb to these thermal stresses in the form of stress cracking.This cracking can be accelerated by high frequency autoclave cycles and/or the use of harsh chemicals, however, it is recognized these outside contributing factors act as a catalyst to the propagation of the cracks which ultimately can lead to catastrophic failure of the shims.Due to crack propagation being recognized as an inherent risk of the shims after repeated use, the potential cause is traced to end of life.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the attune conv shim sz5 14mm has would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot - the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Corrected: h3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CONV SHIM SZ5 14MM
Common Device Name
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18474756
Report Number1818910-2024-00617
Device Sequence Number6944135
Product Code HWT
UDI-Device Identifier10603295135074
UDI-Public10603295135074
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254500930
Device Lot NumberBFA0WS9
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2024
Supplement Date Received by Manufacturer02/07/2024
03/06/2024
03/14/2024
03/26/2024
Initial Report FDA Received Date01/09/2024
Supplement Report FDA Received Date02/07/2024
03/08/2024
03/14/2024
03/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/03/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-