• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Catalog Number 21AGFN-756
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on 15 december 2023 a 21mm regent mechanical heart valve was chosen for implantation.After implant, the leaflets did not open.The valve was explanted and a non-abbott valve was implanted.
 
Additional Manufacturer Narrative
An event of valve leaflets would not open or close and explant was reported.The investigation at abbott revealed orifice was chipped.Morphological and hydrodynamic examination indicated the valve met abbott specifications.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Hydrodynamic testing upon return to abbott and at the time of manufacturing indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18475472
Report Number2135147-2024-00149
Device Sequence Number20556722
Product Code LWQ
UDI-Device Identifier05414734005975
UDI-Public05414734005975
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number21AGFN-756
Device Lot NumberC00015521
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/15/2023
Supplement Date Received by Manufacturer02/01/2024
Initial Report FDA Received Date01/09/2024
Supplement Report FDA Received Date02/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-