| Model Number |
0574-0127 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported that during a patient procedure, using a glidescope titanium lopro t4 reusable laryngoscope, no image was displayed during intubation.The procedure was completed using a backup device which was made available in an unspecified amount of time.No delay in the procedure or harm to the patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer declined the option to have their glidescope titanium lopro t4 reusable laryngoscope returned to verathon for evaluation.Since the device was not returned to verathon, the cause could not be determined.Review of complaint history for the reported serial number (b)(6) did not identify any previous complaints reported to verathon.Trending analysis for the glidescope titanium lopro t4 reusable laryngoscopes does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|
| |
|